Verlagslink DOI: 10.1177/0954411916648717
Titel: Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses
Sprache: English
Autor/Autorin: Haschke, Henning 
Jauch-Matt, Sabrina Y. 
Sellenschloh, Kay 
Huber, Gerd 
Morlock, Michael 
Schlagwörter: modular prosthesis;assembly force;relative motion;stem–neck interface;taper angle difference;arthroplasty;biomechanical phenomena;humans;materials testing;motion;prosthesis design;prosthesis failure;mechanical stress;titanium;vitallium;hip prosthesis;hip replacement
Erscheinungsdatum: Jul-2016
Verlag: Sage Publ.
Quellenangabe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 7 (230): pp. 690-699 (2016)
Zeitschrift oder Schriftenreihe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 
Zusammenfassung (englisch): Bi-modular hip arthroplasty prostheses allow adaptation to the individual patient anatomy and the combination of different materials but introduce an additional interface, which was related lately to current clinical issues. Relative motion at the additional taper interface might increase the overall risk of fretting, corrosion, metallic debris and early failure. The aim of this study was to investigate whether the assembly force influences the relative motion and seating behaviour at the stem-neck interface of a bi-modular hip prosthesis (Metha(®); Aesculap AG, Tuttlingen, Germany) and whether this relation is influenced by the taper angle difference between male and female taper angles. Neck adapters made of titanium (Ti6Al4V) and CoCr (CoCr29Mo) were assembled with a titanium stem using varying assembly forces and mechanically loaded. A contactless eddy current measurement system was used to record the relative motion between prosthesis stem and neck adapter. Higher relative motion was observed for Ti neck adapters compared to the CoCr ones (p < 0.001). Higher assembly forces caused increased seating distances (p < 0.001) and led to significantly reduced relative motion (p = 0.019). Independent of neck material type, prostheses with larger taper angle difference between male and female taper angles exhibited decreased relative motion (p < 0.001). Surgeons should carefully use assembly forces above 4 kN to decrease the amount of relative motion within the taper interface. Maximum assembly forces, however, should be limited to prevent periprosthetic fractures. Manufacturers should optimize taper angle differences to increase the resistance against relative motion.
URI: http://tubdok.tub.tuhh.de/handle/11420/1553
DOI: 10.15480/882.1550
ISSN: 2041-3033
Institut: Biomechanik M-3 
Dokumenttyp: (wissenschaftlicher) Artikel
Lizenzhinweis: Dieser Beitrag ist mit Zustimmung des Rechteinhabers aufgrund einer (DFG geförderten) Allianz- bzw. Nationallizenz frei zugänglich. This publication is with permission of the rights owner freely accessible due to an Alliance licence and a national licence (funded by the DFG, German Research Foundation) respectively.
Enthalten in den Sammlungen:tub.dok

Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat 
0954411916648717.pdfVerlags-PDF3,15 MBAdobe PDFMiniaturbild
Öffnen/Anzeigen
Zur Langanzeige

Seitenansichten

53
Letzte Woche
0
Letzten Monat
23
checked on 21.04.2018

Download(s)

23
checked on 21.04.2018

Google ScholarTM

Prüfe

Altmetric


Alle Ressourcen in diesem Repository sind urheberrechtlich geschützt.