Titel: Investigation on stick phenomena in metal-on-metal hip joints after resting periods
Sprache: English
Autor/Autorin: Wimmer, M. A. 
Nassutt, Roman 
Sprecher, Christoph 
Loos, J. 
Täger, Georg 
Fischer, A. 
Schlagwörter: Biocompatible Materials;Equipment Failure Analysis;Friction;Lubrication;Materials Testing;Particle Size;Prosthesis Design;Surface Properties;Time Factors;Vitallium;Hip Prosthesis;Prosthesis Failure
Erscheinungsdatum: Feb-2006
Verlag: Sage Publications
Quellenangabe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 2 (220): 219-227 (2006-02)
Zeitschrift oder Schriftenreihe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 
Zusammenfassung (englisch): Insufficient understanding of tribological behaviour in total joint arthroplasty is considered as one of the reasons for prosthesis failure. Contrary to the continuous motion input profiles of hip simulators, human locomotion contains motion interruptions. These occurring resting periods can cause stick phenomena in metal-on-metal hip joints. The aim of the present study was to investigate the tribological sensitivity of all-metal bearings to motion interruptions on in vitro test specimens and retrieved implants. Friction and wear with and without resting periods were quantified. Unlike the metal-on-polyethylene joints, the static friction of metal-on-metal joints increased up to micros = 0.3 with rest, while wear appeared to be unaffected. This effect is caused by the interlocking of firmly adhered carbon layers, which were generated from the protein-containing lubricant through tribochemical reactions. Since more than 80 per cent of the retrieved implants exhibited macroscopically visible carbon layers, the increase in friction presumably also occurs under physiological conditions, which is then transferred to the bone-implant interface. These recurrent tangential stress peaks should be considered for the design features of the cup-bone interface, in particular when larger-sized implant heads are used.
URI: http://tubdok.tub.tuhh.de/handle/11420/1593
DOI: 10.15480/882.1590
ISSN: 2041-3033
Institut: Biomechanik M-3 
Dokumenttyp: (wissenschaftlicher) Artikel
Lizenzhinweis: Dieser Beitrag ist mit Zustimmung des Rechteinhabers aufgrund einer (DFG geförderten) Allianz- bzw. Nationallizenz frei zugänglich. This publication is with permission of the rights owner freely accessible due to an Alliance licence and a national licence (funded by the DFG, German Research Foundation) respectively.
Enthalten in den Sammlungen:tub.dok

Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat 
09544119jeim117.pdfVerlags-PDF214,69 kBAdobe PDFMiniaturbild
Öffnen/Anzeigen
Zur Langanzeige

Seitenansichten

33
Letzte Woche
0
Letzten Monat
checked on 21.04.2018

Download(s)

12
checked on 21.04.2018

Google ScholarTM

Prüfe

Altmetric


Alle Ressourcen in diesem Repository sind urheberrechtlich geschützt.