Titel: Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulation
Sprache: English
Autor/Autorin: Jin, Zhongmin 
Meakins, Steven 
Morlock, Michael 
Parsons, Peter A. 
Hardaker, Catherine 
Flett, Magnus 
Isaac, Graham H. 
Schlagwörter: Arthroplasty, Replacement, Hip;Biocompatible Materials;Cadaver;Elasticity;Equipment Design;Equipment Failure Analysis;Hardness;Hardness Tests;Humans;In Vitro Techniques;Materials Testing;Metals;Pressure;Prosthesis Design;Surface Properties;Hip Prosthesis;Prosthesis Failure
Erscheinungsdatum: 1-Feb-2006
Quellenangabe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 2 (220): 299-309 (2006)
Zeitschrift oder Schriftenreihe: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 
Zusammenfassung (englisch): The interference press fit of a metallic one-piece acetabular cup employed for metal-on-metal hip resurfacing procedures was investigated experimentally under laboratory conditions in the present study, in particular regarding the cup deformation. Tests were carried out in cadavers as well as polyurethane foams of various grades with different elastic moduli to represent different cancellous bone qualities. The cadaver test was used to establish the most suitable configuration of the foam model representing realistic support and geometrical conditions at the pelvis. It was found that a spherical cavity, with two identical areas relieved on opposite sides, was capable of creating a two-point pinching action of the ischeal and ilial columns on the cup as the worst-case scenario. Furthermore, the cup deformation produced from such a two-point loading model with a grade 30 foam was similar to that measured from the cadaver test. Therefore, such a protocol was employed in subsequent experimental tests. For a given size of the outside diameter of the cup of 60 mm, the cup deflection was shown to be dependent largely on the cup wall thickness and the diametral interference between cup and prepared cavity at implantation. For a relatively thin cup with a wall thickness between 2.3 mm (equator) and 4 mm (pole) and with a modest nominal diametral interference of 1 mm, which corresponds to an actual interference of approximately 0.5 mm, the maximum diametral cup deflection (at the rim) was around 60 microm, compared with a diametral clearance of 80-120 microm between the femoral head and the acetabular cup, generally required for fluid-film lubrication and tribological performances. Stiffening of the cup, by both thickening and lateralizing by 1 mm, reduced the cup deformation to between 30 and 50 microm with actual diametral interferences between 0.5 and 1 mm.
URI: http://tubdok.tub.tuhh.de/handle/11420/1605
DOI: 10.15480/882.1602
ISSN: 2041-3033
Institut: Biomechanik M-3 
Dokumenttyp: (wissenschaftlicher) Artikel
Lizenzhinweis: Dieser Beitrag ist mit Zustimmung des Rechteinhabers aufgrund einer (DFG geförderten) Allianz- bzw. Nationallizenz frei zugänglich. This publication is with permission of the rights owner freely accessible due to an Alliance licence and a national licence (funded by the DFG, German Research Foundation) respectively.
Enthalten in den Sammlungen:tub.dok

Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat 
095441105x69150.pdfVerlags-PDF323,1 kBAdobe PDFMiniaturbild
Öffnen/Anzeigen
Zur Langanzeige

Seitenansichten

48
Letzte Woche
0
Letzten Monat
1
checked on 19.10.2018

Download(s)

27
checked on 19.10.2018

Google ScholarTM

Prüfe


Alle Ressourcen in diesem Repository sind urheberrechtlich geschützt.